|
Resistive Plate
Chambers |
|
|
RPC Testing
Procedure |
|
|
|
PAEC is responsible for the assembly of RPCs (Resistive Plate Chambers). In order to keep up RPCs' quality during assembly process, following pre-assembly and post-assembly tests are done at PAEC. |
|
|
Gas Gaps Testing |
- Check the dimensions of the gas gaps at random positions
- Note down the buckling of the gaps and reject gaps bent more than
8mm
- Check the gas inlets, HV and ground connections for any damage
- Store gas gaps which pass initial visual test under humidity and temperature control area
|
-
Block two inlets of gas gaps and connect other two inlets to the gas system and manometer respectively
-
Keep the gap under 20-mbar over pressure for 10 minutes and observe reading on the manometer
-
Reject the gap, if the pressure column on the manometer is reduced by 2 cm after 10 minutes
|
- Place a template sheet where each spacer location is marked on the gas gap
-
Note the height of water column in the manometer after 10 minutes of 20-mbar of over pressure in the gas gap
-
Press each spacer twice (5 N force at least) and check the water column of the manometer. If there is a sudden jump of about 1.5 -3 cm then that spacer is a pop spacer
- Note this information in the gas gap test check list
|
- High
Voltage vs. Dark Current Test
- Place gaps under high voltage with the gas mixture (Freon 95% and Iso-butane 4.5%) under 2 mbar pressures at 5 l/h for 8-10 hours
- Measure the dark current. Accept the gaps with less than
5-6 ľA
currents for the assembly
- Note the high voltage and dark current and plot the graphs
- Save the data and plots in the database
- Perform all the tests under humidity and temperature control. Save the temperature and humidity readings in the database
|
|
Other Quality Control Tests |
|
- Test the FEB boards in the testing jig provided by Matrix, Srl , Italy
- Store all the data for each FEB board in the database
|
|
- Coolant Pipe
Pressure Test
-
Block one end of the pipe and connect the other end to the oxygen cylinder with pressure regulator
-
Gradually increase the pressure up to
20-bar
and maintain it for 5 minutes
-
Use soap solution to check for any leakage at joints and brazing ends
|
| After assembly, the chambers are moved from PAEC to NCP for further testing and quality assurance using cosmic telescope. Following tests are performed at NCP.
|
|
- All chambers are inspected visually, refer RPC Quality Assurance Sheet
|
|
Testing using Cosmic Telescope |
|
- Use gas mixture consisting of 96.5% Freon, 3.0 % Iso-butane, and 0.5 % SF 6
- Apply the high voltage after 8 volume changes
- Gradually raise HV to 8.6 kV. Leave
chamber for 6 hours and note the dark currents at regular time
intervals
|
- High
Voltage Scan
- Perform HV scan with the following HV points::
-
8600, 8800, 9000, 9200, 9300, 9400, 9500 & 9600 V
- Take 20,000 events for each HV point. Use offline analysis program to calculate the following chamber parameters:
- Dead and Noisy Strips
- Efficiency
- Cluster Size
- Rate
- Probability of Spurious Hits
- Record all results on a sheet as well as in the database.
-
Reject the chamber, if
the measured efficiency is less than 95% or the cluster
size is bigger than 3 at the nominal operating voltage of
9300 V
|
|
|
RPCs Testing
Status |
|
|
|
 |
|